Chinese Team Successfully Implants EV-ICD in Patient with Severe Thoracic Deformity
November 5, 2026
November 5, 2026
Recently, a team led by Professor Qiming Liu at the Second Xiangya Hospital of Central South University published a case report in European Heart Journal – Case Reports titled “Successful extravascular implantable cardioverter defibrillator implantation in a patient with ankylosing spondylitis and severe thoracic deformity: a case report.”

The patient had severe ankylosing spondylitis, marked thoracic and spinal deformities, and coexisting dilated cardiomyopathy, placing him at risk of sudden cardiac death. Following a comprehensive assessment, the cardiology team at the Second Xiangya Hospital of Central South University selected an extravascular implantable cardioverter defibrillator (EV-ICD). Through individualized patient positioning, meticulous anatomical dissection and multidisciplinary collaboration, the team successfully completed the implantation.
The successful treatment of this particularly challenging case provides an example of how EV-ICD implantation can be approached in patients with complex thoracic anatomy, offering clinical experience from China to the international medical community.
Severe Thoracic Deformity: Which ICD Should Be Selected?
The patient presented with exertional dyspnea and palpitations and was diagnosed with dilated cardiomyopathy. Holter monitoring revealed frequent multifocal premature ventricular complexes (PVCs) and non-sustained ventricular tachycardia. Cardiac magnetic resonance imaging also identified mid-wall late gadolinium enhancement in the interventricular septum, suggesting an underlying arrhythmogenic substrate.
After six months of optimized medical therapy, the patient's left ventricular ejection fraction (LVEF) remained below 35%, and his functional status was New York Heart Association (NYHA) Class II. Considering his cardiac function and arrhythmic risk, the team determined that he met the criteria for a Class I primary-prevention indication for ICD implantation. The main challenge was selecting the appropriate device and implantation approach.
The patient had a long history of ankylosing spondylitis, with severe kyphosis resulting in substantial changes in thoracic anatomy, as well as ossification in the costoclavicular region.
Although a conventional transvenous ICD provides established defibrillation and antitachycardia pacing (ATP) therapies, several factors required careful consideration in this patient, including venous access, potential long-term lead compression and complications associated with intravascular leads.
A subcutaneous ICD (S-ICD) was also considered. However, the markedly altered thoracic anatomy raised questions about the suitability of its sensing vectors. In addition, this technology does not provide conventional ATP therapy.
By comparison, an EV-ICD places its lead in the substernal anterior mediastinum, avoiding transvenous leads while retaining ATP capability. Following multidisciplinary assessment involving cardiac electrophysiology, cardiac surgery, anesthesiology and rheumatology specialists, the team decided to proceed with EV-ICD implantation.
However, this approach presented its own challenges. Preoperative CT revealed that the right ventricle was directly adjacent to the posterior surface of the sternum, leaving virtually no retrosternal space for procedural access. Establishing the substernal tunnel therefore carried a risk of cardiac injury or perforation.
At the same time, the relatively flat contour of the sternum and the distance between the right atrium and right atrial appendage and the anticipated lead trajectory were considered potentially favorable for reducing the risk of P-wave oversensing.
The team developed an individualized surgical plan and discussed potential intraoperative assistance with the cardiac surgery department.

From Patient Positioning to Meticulous Dissection: Completing a Challenging Implantation
The first challenge was determining an appropriate surgical position. Severe kyphosis prevented the patient from lying in a conventional supine position and also affected lateral fluoroscopic imaging.

To accommodate the spinal deformity, the surgical team placed four to five folded surgical gowns beneath the patient's neck and upper back. This adjustment improved neck extension and fluoroscopic visualization, creating more favorable conditions for the subsequent procedure.
The most significant technical challenge arose during the creation of the substernal tunnel.
After general anesthesia was administered, the surgeon made a paraxiphoid incision and performed layer-by-layer dissection into the substernal region. The posterior surface of the sternum and cardiac pulsations could be felt with a finger. However, significant resistance was encountered when advancing further toward the distal end, preventing completion of the dissection.
Because the right ventricle was positioned directly behind the sternum, forceful advancement could have increased the risk of cardiac injury.
The electrophysiology team consulted with cardiac surgeons during the procedure. After reassessing the local anatomy, the team carefully identified and dissected the surrounding tissue to confirm the correct substernal anatomical plane. The tunneling tool was then advanced, and the defibrillation lead was delivered through the sheath to the intended position.

Both anteroposterior and lateral fluoroscopic views confirmed that the lead followed the posterior surface of the sternum and was positioned within the intended anatomical region.
The unusual resistance encountered during substernal dissection may have been associated with inflammatory adhesions related to ankylosing spondylitis, although the restricted retrosternal space may also have contributed.
Preoperative CT can help physicians identify anatomical risks in patients with complex deformities, but intraoperative decisions must still be guided by actual surgical findings. In this case, individualized positioning and careful dissection enabled successful implantation, highlighting the importance of tailored procedural strategies in challenging EV-ICD cases.
Successful Defibrillation with a Single 30 J Shock, with Stable Six-Month Follow-Up
Following lead implantation, the team assessed the EV-ICD system's sensing, pacing and defibrillation performance.
The intraoperative R-wave sensing amplitude was 5.6 mV, with no P-wave oversensing. High-voltage impedance was 95 Ω, while low-voltage impedance was 646 Ω.

During defibrillation threshold testing, the device successfully detected induced ventricular fibrillation and restored sinus rhythm with a single 30 J shock. Compared with the system's maximum output of 40 J, this represented a 10 J defibrillation safety margin.
The patient was discharged without complications. During the six-month follow-up period, device sensing and pacing parameters remained stable, with no sensing abnormalities or inappropriate shocks reported.
Despite severe thoracic deformity and an extremely restricted retrosternal space, the lead was successfully positioned as intended. Intraoperative defibrillation testing was successful, and device parameters remained stable during the six-month follow-up.
Chinese Experts Share Clinical Experience with EV-ICD Implantation in Complex Cases
As EV-ICD technology enters clinical practice, physicians must address not only cardiac anatomical challenges but also abnormal vascular access and unusual thoracic structures when selecting an appropriate lead implantation route.
The experience gained by this Chinese clinical team highlights the importance of individualized decision-making throughout the implantation pathway.
In the presence of severe thoracic deformity, the physicians compared three approaches: transvenous ICD, subcutaneous ICD and EV-ICD. Their selection was based on the patient's specific anatomical characteristics and clinical needs.
When the retrosternal space was almost completely absent, the team used CT imaging to assess the spatial relationship between the heart and sternum and carefully confirmed the anatomical plane for tunnel creation.
From device selection and preoperative planning to intraoperative management, every stage required careful consideration of the patient's unusual anatomy.
For physicians considering EV-ICD implantation in patients with complex anatomical conditions, thorough preoperative risk assessment is essential. Individualized plans should address patient positioning, fluoroscopic imaging and the lead dissection pathway.
The clinical adoption of an innovative medical technology depends on accumulating practical experience across different patient populations and anatomical conditions. The extravascular lead placement and ATP capability of EV-ICD offer an additional treatment option for selected patients. However, physicians must still address several practical questions: whether the patient's thoracic anatomy permits lead implantation, whether sensing and defibrillation parameters meet clinical requirements, how intraoperative difficulties should be managed, and whether device performance remains stable during follow-up.
Conclusion
From preoperative CT revealing an almost completely obliterated retrosternal space to meticulous surgical dissection, defibrillation threshold testing and structured follow-up, this case report documents how a Chinese clinical team addressed the challenges of EV-ICD implantation.
The team's experience with treatment decisions, anatomical assessment and intraoperative management in a complex patient provides a valuable contribution to international clinical discussions.
As EV-ICD technology is adopted by more clinical centers worldwide, experience from Chinese specialists may help inform treatment decisions and implantation strategies for patients with unusual thoracic deformities, while providing useful clinical reference for the further adoption and application of EV-ICD therapy.
Expert Profile

Qiming Liu, MD, PhD, is a First-Class Chief Physician, Second-Level Professor and doctoral supervisor. He is the inaugural Director of the Department of Cardiovascular Medicine at Xiangya School of Medicine, Central South University, Vice President of the Cardiovascular Hospital of the Second Xiangya Hospital of Central South University, and Director of the Department of Cardiovascular Medicine at the Second Xiangya Hospital.
He is recognized as a Hunan Provincial Furong Program High-Level Health Talent. His professional affiliations include Fellow of the American College of Cardiology (FACC) and Fellow of the Heart Rhythm Society (FHRS). He also serves as a Standing Committee Member of the Chinese Society of Pacing and Electrophysiology, a Committee Member of the Chinese Society of Cardiology, a Committee Member of the Chinese Medical Doctor Association's Integrative Medicine Branch, and Deputy Director of the Arrhythmia and Electrocardiology Branch of the China International Exchange and Promotive Association for Medical and Health Care. In addition, he is Director of the Hunan Provincial Quality Control Center for Cardiovascular Interventional Therapy and an expert at the National Medical Quality Control Center for Arrhythmia Interventional Technology.
He has undertaken visiting academic appointments at the Heart Rhythm Institute of the University of Oklahoma in the United States, St. Georg Hospital in Hamburg, Germany, and the Cardiac Electrophysiology Center of the University of Montreal in Canada. He also serves on the editorial boards of several domestic and international medical journals, including PACE and the Chinese Journal of Arrhythmias.
In recent years, he has published more than 100 research papers and academic articles as first or corresponding author in leading international journals, including The New England Journal of Medicine, The Lancet, Journal of the American College of Cardiology, International Journal of Cardiology and Circulation Journal.
He has led five General Program projects funded by the National Natural Science Foundation of China, three projects funded by the Hunan Provincial Natural Science Foundation, two key projects of the Ministry of Science and Technology of China, and two major research projects funded by Hunan Province.
He is the chief editor of China's first monograph on leadless cardiac pacemakers, Leadless Cardiac Pacemakers, and the national key publication under China's 11th Five-Year Plan, Common Problems and Solutions in Cardiac Catheterization Procedures. He has also contributed to the cardiology textbook Internal Medicine, published by Higher Education Press. His honors include the "Health Guardian" title awarded by China's National Health Commission and recognition as one of the "Top Ten Medical Innovation Experts" in 2021.